FST Status Messages

 

General Display Messages
OFF Instrument has been powered OFF
bAt Indicates battery charge is low
bLo Indicates that the instrument is waiting for a sample to be provided
Flo/Lo Indicates that during the provided sample flow rates below the minimum allowed flow rate were detected
Flo/Hi Indicates that the provide sample had flow rates greater than the maximum allowed flow rate
Flo/InS Indicates that the flow rate was too inconsistant to be considered a valid sample
Flo/Cut Indicates that the sample was abruptly stopped at the end of the blow resulting in an invalid sample
POS This is a non-digital result indicating that the measured alcohol concentration was at or above the POSITIVE cutoff
NEG This is a non-digital result indicating that the measured alcohol concentration was below the POSITIVE cutoff
PAS This is a non-digital result indicating that the measured alcohol concentration was below the CAUTION cutoff
CAU This is a non-digital result indicating that the measured alcohol concentration was below the FAIL cutoff
Fail This is a non-digital result indicating that the measured alcohol concentration was equal to or above the FAIL cutoff
Status Messages
E03 Sensor Blank Timeout (> 60 seconds).  Too long to establish Fuel Cell Baseline.
E06 Insufficient Sample.  Subject unable to provide adequate sample on preset number of allowed sample attempts – normally 3 attempts.
E07 Calibration Failed (unable to calibrate – usually due to a sample out of range). Too low/high of Fuel Cell output difference from last calibration.
E09 Temp too Cold for Test Type Being Performed.
E10 Temp too Hot for Test Type Being Performed.
E11 Failed Air Blank – Replace mouthpiece with new mouthpiece and begin test again.
E12 Invalid Sample – instrument detected an anomaly in the fuel cell wave form during analysis and aborted test sequence. Could be RFI
E21 Fuel Cell leak detected
E22 Invalid Calibration Target (less than .040 or greater than .400
E23 Memory Failure
E24 Suck back detected
E25 Voltage low on Reset Solenooid
E26 Excessive voltage on Reset Solenoid
E27 Insufficient drive voltage to sample.
E28 Excessive Voltage After Sample
E29 Fuel Cell Signal Clipping Condition – Fuel Cell reading exceeded A/D maximum range.
E30 Pressure Sensor Baseline Error
E31 Result Over Range – Fuel cell result greater than configured maximum reading.
E32 Internal Error – normally hardware or firmware related.
Settings/Options Menu
rcL Recall last test result
PAS Passive Testing option
rbL Road block or Quick Test option
ACC Accuracy Check
CAL Calibration Adjustment
dSr Display Software Revision identifier